The vapour pressure of solid cadmium
TA O'Donnell
Australian Journal of Chemistry
8(4) 485 - 492
Published: 1955
Abstract
An apparatus has been designed for measurement by the Knudsen effusion method of the vapour pressure of solid cadmium in the temperature range 200-260 °C. Reliability of the apparatus was checked by measurement of the vapour pressure of zinc, for which a recent, reliable determination has been reported. The vapour pressure of cadmium in the temperature range studied is given by the equation log10 Pmm = 9.413 ? 6078/T .https://doi.org/10.1071/CH9550485
© CSIRO 1955