Pilot of non-invasive (oral fluid) testing for HIV within a clinical setting
Joseph Debattista A C , Gregory Bryson B , Nicole Roudenko B , John Dwyer A , Mark Kelly A , Patrick Hogan B and John Patten AA Brisbane Sexual Health and AIDS Service, Northside Health Service District, Brisbane, Qld 4000, Australia.
B Division of Immunology, Queensland Health Pathology Services, Royal Brisbane and Women’s Hospital, Herston, Qld 4029, Australia.
C Corresponding author. Email: Joseph_Debattista@health.qld.gov.au
Sexual Health 4(2) 105-109 https://doi.org/10.1071/SH07014
Submitted: 7 March 2007 Accepted: 30 April 2007 Published: 23 May 2007
Abstract
Background: The objectives of the present study were: to determine the sensitivity and specificity of oral fluid testing compared with the performance of standard blood-based HIV enzyme immunoassay; to assess the feasibility of oral fluid specimen collection from clients for the purposes of HIV testing within a clinical setting; and to assess the clinical and laboratory impact regarding staffing, material resources, expertise and funding of oral fluid testing. Methods: A parallel comparative trial of oral fluid and blood testing was conducted among a group of HIV positive clients and a group of unknown HIV serostatus clients where each client was offered both tests. An ambulatory HIV clinic recruited 175 known HIV positive clients and 179 persons were recruited through an inner city sexual health clinic while attending for routine sexual health checks. Client responses to oral fluid collection were assessed. The sensitivity and specificity of oral fluid testing were calculated. Results: Of the 176 confirmed HIV reactive blood test results, the OraSure (OraSure Technologies, Beaverton, OR, USA) assay failed to detect only one of these, demonstrating a sensitivity of 99.4%. Of the 178 blood specimens that were tested as non-reactive by the AxSYM (Abbott Laboratories, Abbott Park, IL, USA) Combo system, OraSure recorded four of the corresponding oral fluid specimens as reactive (assumed to be false-positive), giving a specificity of 97.6%. Although evaluation of patients undergoing the test showed a large proportion (88.6%) preferred the OraSure test to conventional blood testing, a large minority of these (22.6%) made such a preference conditional on the OraSure test being as reliable as current blood testing. Conclusions: This limited clinic based trial of oral fluid testing for HIV antibodies among an outpatient population has demonstrated the potential of oral fluid as a specimen for HIV testing. However, the lower performance of the test compared with current serum-based tests may limit the usefulness of OraSure to epidemiological studies or as an alternative screening tool in outreach settings among higher risk populations.
Acknowledgements
We gratefully acknowledge the support of the laboratory staff of the Division of Immunology, QHPS who performed the assays, the clinical staff of the Brisbane Sexual Health and AIDS Service who collected the specimens and patient surveys, Garry Boddy of the Communicable Diseases Unit, Dr Harry Bartlett of the Queensland University of Technology for his statistical advice, and most importantly, patients who generously participated. This study was funded by the Communicable Diseases Unit, Queensland Health.
[1] Janssen RS, Holtgrave DR, Valdiserri RO,, Shepherd M, Gayle HD, De Cock KM. The serostatus approach to fighting the HIV epidemic: prevention strategies for infected individuals. Am J Public Health 2001; 91 1019–24.
| PubMed |
[2] Hilton C, Sabundayo B, Langan S, Hilton M, Henson C, Quinn T, et al. Screening for HIV infection in high risk communities by urine antibody testing. J Acquir Immune Defic Syndr 2002; 31 416–21.
| PubMed |
[3] Vajdic CM, Middleton M, Bowden FJ, Fairley CK, Kaldor JM. The prevalence of genital Chlamydia trachomatis in Australia 1997–2004: a systematic review. Sex Health 2005; 2 169–83.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[4] Mylonakis E, Paliou M, Lally M, Flanigan T, Rich J. Laboratory testing for infection with the human immunodeficiency virus: established and novel approaches. Am J Med 2000; 109 568–76.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[5] Gallo D, George JR, Fitchen JH, Goldstein AS, Hindahl MS. Evaluation of a system using oral mucosal transudate for HIV-1 antibody screening and confirmatory testing. OraSure HIV Clinical Trials Group. JAMA 1997; 277 254–8.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[6] Malamud D. Oral diagnostic testing for detecting human immunodeficiency virus-1 antibodies: a technology whose time has come. Am J Med 1997; 102 9–14.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[7]
[8] Meehan M, Sewankambo N, Wawer M, McNairn D, Quinn T, Lutalo T, et al. Sensitivity and specificity of HIV testing of urine compared with sperum specimens: Rakai Uganda. Sex Transm Dis 1999; 26 590–2.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[9] Dodds JP, Mercey DE, Parry JV, Johnson AM. Increasing risk behaviour and high levels of undiagnosed HIV infection in a community sample of homosexual men. Sex Transm Infect 2004; 80 236–40.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[10] Vargo S, Agronick G, O’Donnell L, Stueve A. Using peer recruitment and OraSure to increase HIV testing. Am J Public Health 2004; 94 29–31.
| PubMed |
[11] Osmond DH, Catania J, Pollack L, Canchola J, Jaffe D, MacKellar D, et al. Obtaining HIV test results with a home collection test kit in a community telephone sample. J Acquir Immune Defic Syndr 2000; 24 363–8.
| PubMed |
[12] Sy FS, Rhodes SD, Choi ST, Drociuk D, Laurent AA, Naccash RM, et al. The acceptability of oral fluid testing for HIV antibodies. A pilot study in gay bars in a predominantly rural state. Sex Transm Dis 1998; 25 211–5.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[13] Granade TC, Phillips SK, Parekh B, Gomez P, Kitson-Piggott W, Oleander H, et al. Detection of antibodies to human immunodeficiency virus type 1 in oral fluids: a large-scale evaluation of immunoassay performance. Clin Diagn Lab Immunol 1998; 5 171–5.
| PubMed |
[14] Emmons W. Accuracy of oral specimen testing for human immunodeficiency virus. Am J Med 1997; 102 15–20.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[15] Pugatch D, Levesque B, Greene S, Strong L, Flanigan T, Lally M. HIV testing in the setting of inpatient acute substance abuse treatment. Am J Drug Alcohol Abuse 2001; 27 491–9.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[16] Pugatch DL, Levesque BG, Lally MA, Reinert SE, Filippone WJ, Combs CM, et al. HIV testing among young adults and older adolescents in the setting of acute substance abuse treatment. J Acquir Immune Defic Syndr 2001; 27 135–42.
| PubMed |
[17] Clair S, Singer M, Huertas E, Weeks M. Unintended consequences of using an oral HIV test on HIV knowledge. AIDS Care 2003; 15 575–80.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[18] Summers T, Spielberg F, Collins C, Coates T. Voluntary counselling, testing and referral for HIV: new technologies, research findings create dynamic opportunities. J Acquir Immune Defic Syndr 2000; 25 S128–35.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[19] George JR, Fitchen JH. Future applications of oral fluid specimen technology. Am J Med 1997; 102 21–5.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[20] Judd A, Parry J, Hickman M, McDonald T, Jordan L, Lewis K, et al. Evaluation of a modified commercial assay in detecting antibody to hepatitis C virus in oral fluids and dried blood spots. J Med Virol 2003; 71 49–55.
| Crossref | GoogleScholarGoogle Scholar | PubMed |
[21] Judd A, Hutchinson S, Wadd S, Hickman M, Taylor A, Jone S, et al. Prevalence of, and risk factors for hepatitis C virus infection among recent initiates to injecting in London and Glasgow: cross sectional analysis. J Viral Hepat 2005; 12 655–62.
| Crossref | GoogleScholarGoogle Scholar | PubMed |